Polytron MUI65 Series AC-DC Converters, 65 Watt Medical Universal Input Range
|
Product Options
Daburn Part Number: |
|
|
With a universal input range and high efficiency up to 92%, our 65-watt MUI65 series of AC-DC converters for medical applications has an operating temperature range of -25 to 80°C, low leakage current under 75µA and a built-in EMI class A filter. This model also features an operating altitude of 5000M, 0.11 watt low standby power consumption and a 2000 VAC input to output 2 MOPP isolation. In addition, this model meets all UL, CE and ANSI/AAMI ES60602-2, EN60601-1 and IEC60601-1 safety standards.
FEATURES: |
• Universal Input Range • Operating Temperature Range: -25º to 80ºC (with derating) • Compact Package: 2.00” × 3.00” × 1.16” • Mounting Options: Screw Terminal, Chassis Mount, Din Rail, Open Type, Enclosed (Standard)
|
• High Efficiency: Up to 92% • Active Power Factor Correction • Low Leakage Current Under 75μA • Built-in EMI Class A Filter • 2000VAC Input to Output 2 MOPP Isolation
|
• Class I and Class II Protection • Medical Safety Approvals: UL, CE and ANSI/AAMI ES60601-1, EN60601-1 and IEC60601-1 3rd Edition • RoHS Compliant to 2011/65/EU • 0.11 Watt Low Standby Power Consumption • Operating Altitude 5000M
|
MUI65 Series Specification Sheet PDF
Daburn Part Number |
Input Range Vac |
Output Voltage Vdc |
Output Current at Convention Cooled 60ºC TA A |
Max. Output Power W |
Input Power at No Load W |
Efficiency (%) |
MUI65-5SB |
85-264 |
5 |
10 |
50 |
0.11 |
90 |
MUI65-7SB |
7.5 |
8.67 |
65 |
0.11 |
90 |
MUI65-9SB |
9 |
7.23 |
65 |
0.11 |
91 |
MUI65-12SB |
12 |
5.42 |
65 |
0.11 |
92.5 |
MUI65-15SB |
15 |
4.34 |
65 |
0.11 |
93.5 |
MUI65-24SHB |
24 |
2.71 |
65 |
0.11 |
93.5 |
MUI65-24SB |
24 |
2.71 |
65 |
0.11 |
92 |
MUI65-28SHB |
28 |
2.33 |
65 |
0.11 |
93.5 |
MUI65-28SB |
28 |
2.33 |
65 |
0.11 |
91.5 |
MUI65-36SB |
36 |
1.81 |
65 |
0.11 |
92.5 |
MUI65-48SB |
48 |
1.36 |
65 |
0.11 |
93 |
MUI65-53SB |
53 |
1.24 |
65 |
0.11 |
92.5 |
Input and Output Specifications:
|
INPUT
- Voltage Range
- AC input: 85 - 264 Vac
- DC input: 120 - 370 Vdc
- Input frequency
- Input current
- 100 Vac and full load: 1.6 A
- 240 Vac and full load: 0.9 A
- No load input power
- Leakage current
- Start-up time: 1000 ms
- Rise time: 20 ms
- Hold-up time
- 115 Vac and full load: 16 ms
- Input inrush current
- Input protection
- Internal fuse in line and neutral: T3.15A/250Vac
|
OUTPUT
- Output power: 65 Watts
- Initial set voltage
- 230 Vac and full load: ±1%
- Line regulation
- Low line to high line at full load: ±0.2%
- Load regulation
- No load to full load, 5Vout: ±0.7%
- Output voltages: ±0.5%
- 10% load to 90% load, 5Vout: ±0.6%
- 10% load to 90% load, others: ±0.4%
- Voltage adjustability
- Single output, 53Vout: -20 - 10%
- Single output, others: ±10%
- Minimum load: 0%
- Ripple and noise (measured by 20 MHz bandwidth)
- With a 10µF/25V 1206 X7R MLCC, 5Vout, 7.5Vout, 9Vout, 12Vout, 15Vout: 75 mVp-p
- With a 1µF/50V 1206 X7R MLCC, 24Vout, 28Vout, 36Vout: 75 mVp-p
- With a 0.1µF/100V 1206 X7R MLCC, 48Vout, 53Vout: 150 mVp-p
- Temperature coefficient: ±0.02%/°C
- Transient response peak deviation
- Load step change from 50-75% at 2.5A/µs: 3Vout
- Transient response recovery time: 600µs
- Over voltage protection
- % of Vout(nom); latch mode: 125 - 140%
- Over load protection
- % of Lout rated; hiccup mode: 145%
- Short circuit protection: continuous, automatic recovery
|
More Specifications:
|
GENERAL SPECIFICATIONS
- Isolation voltage (1 min, 2 MOPP isolation)
- Input to Output: 4000 Vac
- Input (output) to FG: 2500 Vac
- Isolation resistance: 500 Vdc: 0.1 GΩ
- Switching frequency
- 5Vout: 230 Vac, 60 kHz
- 7.5Vout: 80 kHz
- 9Vout: 70 kHz
- Others: 120 kHz
ENVIRONMENTAL SPECIFICATIONS
- Operating ambient temperature
- Natural convection with derating: -40 - 80°C
- Storage temperature range: -40 - 85°C
- Operating altitude
- Vertical direction by natural convection (20LFM): 5000 m
- Shock: IEC60068-2-27
- Vibration: IEC60068-2-6
- Relative humidity
- Non-condensing: 5 to 95% RH
PHYSICAL SPECIFICATIONS
- Design meet safety standard
- ANSI/AAMI, ES60601-1, IEC60601-1, EN60601-1
- Weight
- MUIO: 117g (4.13oz)
- MUIT: 157g (5.54oz)
- MUI: 172g (6.07oz)
- MUID: 193 (6.81oz)
- Dimensions: 2.00" x 3.00" x 1.16" (50.8mm x 76.2mm x 29.46mm)
- MTBF: 1.494 x 10^6 hrs, MIL-HDBK-217F, full load
|
EMC SPECIFICATIONS
- EMI
- EN55011, EN55022 and FCC Part 18
- Conducted: Class B
- Radiated: Class B
- Harmonic currents
- EN6100-3-2, full load: Class A
- Voltage flicker: EN61000-3-3
- ESD
- EN61000-4-2, Perf. Criteria A
- Air ±8KV and Contact ±6KV
- Radiated immunity
- EN61000-4-3, Perf. Criteria A: 20 V/m
- Fast transient
- EN61000-4-4, Perf. Criteria A: ±2KV
- Surge
- EN61000-4-5, Perf. Criteria A
- Conducted immunity
- EN61000-4-6, Perf. Criteria A: 20 Vr.m.s
- Power frequency magnetic field
- EN6100-4-8, Perf. Criteria A: 10 A/M
- Dip and interruptions
- EN60601-1-2
- 230 Vac 50Hz, 30%, 500mS, Perf. Criteria A
- 230 Vac, 50Hz, 60%, 100mS, Perf. Criteria A
- 230 Vac, 50Hz, >95%, 10mS, Perf. Criteria A
- 230 Vac, 50Hz, >95%, 5000mS, Perf. Criteria B
- EN61000-4-11
- 100 Vac, 50Hz, 30%, 500mS, Perf. Criteria A
- 100 Vac, 50Hz, 60%, 100mS, Perf. Criteria B
- 100 Vac, 50Hz, >95%, 10 mS, Perf. Criteria A
- 100 Vac, 50Hz, >95%, 5000mS, Perf. Criteria B
|
|
With an internal input filter, our 10 watt single and dual output UIL10 AC-DC power supply module is CE certified and features short circuit, overload and over voltage protection.
|
With short circuit and overload protection, our UIL5 series of AC-DC converters feature high efficiency up to 80%, universal input of 90-264 Vac and a 2 year warranty. This series also meets all UL60950-1 and IEC60950-1 safety standards.
|
Designed for all Single and Dual Linear PC/Board Mount
Power Supplies in Cases A, B, C, D (see specification sheet PDF)
|