Liquid Ring QT Pump Specifications
|
|
Pump Horsepower
|
Capacity (SCFM) @ 19"Hg
|
System Btu/hr
|
System dB(A) Level
|
System FLA
|
|
Pump
|
System
|
208V
|
230V
|
460V
|
380V (50 Hz)
|
|
Duplex
|
|
3
|
14
|
14
|
7,635
|
73
|
18.9
|
17.4
|
10.7
|
8.8
|
|
5
|
21
|
21
|
12,725
|
76
|
31.2
|
28.2
|
15.9
|
18.3
|
|
5 [HC]
|
26
|
26
|
12,725
|
72
|
31.0
|
28.0
|
14.3
|
17.7
|
|
7.5
|
38
|
38
|
19,088
|
73
|
44.5
|
40.2
|
24.5
|
26.2
|
|
10
|
53
|
53
|
25,450
|
77
|
54.9
|
49.6
|
26.3
|
31.4
|
|
15
|
76
|
76
|
38,175
|
76
|
95.8
|
91.8
|
45.9
|
52.9
|
|
20
|
110
|
110
|
50,900
|
76
|
127.1
|
122.2
|
61.1
|
66.0
|
|
30
|
172
|
172
|
76,350
|
77
|
196.1
|
177.4
|
88.5
|
89.6
|
|
Triplex
|
|
3
|
14
|
27
|
15,270
|
75
|
28.3
|
26.0
|
16.0
|
13.2
|
|
5
|
21
|
42
|
25,450
|
78
|
46.8
|
42.3
|
23.8
|
27.5
|
|
5 [HC]
|
26
|
52
|
25,450
|
74
|
46.5
|
42.0
|
21.4
|
26.6
|
|
7.5
|
38
|
76
|
38,175
|
75
|
66.7
|
60.3
|
36.8
|
39.3
|
|
10
|
53
|
106
|
50,900
|
79
|
82.3
|
74.4
|
39.5
|
47.1
|
|
15
|
76
|
152
|
76,350
|
78
|
143.7
|
137.8
|
68.9
|
79.3
|
|
20
|
110
|
220
|
101,800
|
78
|
190.6
|
183.3
|
91.7
|
99.0
|
|
30
|
172
|
344
|
152,700
|
79
|
294.1
|
266.1
|
132.8
|
134.4
|
|
Quadraplex
|
|
3
|
14
|
41
|
22,905
|
76
|
37.7
|
34.7
|
21.4
|
17.6
|
|
5
|
21
|
63
|
38,175
|
79
|
62.4
|
56.3
|
31.8
|
36.6
|
|
5 [HC]
|
26
|
78
|
38,175
|
75
|
62.0
|
55.9
|
28.6
|
35.4
|
|
7.5
|
38
|
114
|
57,263
|
76
|
89.0
|
80.4
|
49.0
|
52.4
|
|
10
|
53
|
159
|
76,350
|
80
|
109.8
|
99.2
|
52.6
|
62.8
|
|
15
|
76
|
228
|
114,525
|
79
|
191.6
|
183.7
|
91.8
|
105.7
|
|
20
|
110
|
330
|
152,700
|
79
|
254.2
|
244.4
|
122.2
|
132.0
|
|
30
|
172
|
516
|
229,050
|
80
|
392.2
|
354.8
|
177.0
|
179.2
|
|
Notes:
|
|
• System capacity shown is with one pump in reserve, per NFPA 99
|
|
• All values assume 60Hz motors
|
|
• To convert free air capacity (SCFM) to expanded air capacity (ACFM) at 19”Hg, multiply SCFM by 2.74
|
|
• Maximum recommended ambient temperature: 95°F for standard systems
|
|
• Maximum recommended water temperature: 95°F for standard systems
|
Liquid Ring Pump Fresh Water Usage (Partial Recirculation)
|
Horsepower
|
Water Usage (gpm)
|
|
3
|
1.0
|
|
5
|
2.0
|
|
5 [HC]
|
4.0
|
|
7.5
|
3.5
|
|
10
|
3.5
|
|
DataSheet
|
Description
|
|
1DQT(5-30)VM120
|
Duplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 120 Gallon Vacuum Tank.
|
|
1DQT(5-30)VM200
|
Duplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 200 Gallon Vacuum Tank.
|
|
1DQT(5-30)VM240
|
Duplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 240 Gallon Vacuum Tank.
|
|
1TQT(5-30)VM120
|
Triplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 120 Gallon Vacuum Tank.
|
|
1TQT(5-30)VM200
|
Triplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 200 Gallon Vacuum Tank.
|
|
1TQT(5-30)VM240
|
Triplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 5-30 Horsepower Pumps, 240 Gallon Vacuum Tank.
|
|
1QQT(10-30)VM200
|
Quadruplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 10-30 Horsepower Pumps, 200 Gallon Vacuum Tank.
|
|
1QQT(10-30)VM240
|
Quadruplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 10-30 Horsepower Pumps, 240 Gallon Vacuum Tank.
|
|
1QQT(10-30)VM400
|
Quadruplex, Liquid Ring Pumps - Once Through, Modular with Vertical Tank Mount. 10-30 Horsepower Pumps, 400 Gallon Vacuum Tank.
|
|
Product Reviews
|
|