MUI50 Series Medical AC-DC Power Supplies
|
Price (Each): Call or e-mail for quotes or to order.
Available to Order
Item Number: MUI50
|
Medical Applications
MUI50 Series AC-DC power supplies for medical applications deliver up to 50 watts of output power and up to 70 watts peak power for 5 seconds (typ.), while accepting a universal input of 85 to 264V AC. The Isolation Class I and Class II units provide 4,000V AC input to output 2xMOPP insulation for patient safety and satisfy IEC/EN/ANSI/AAMI ES and IEC/EN/UL 62368-1 safety standards. They feature an adjustable output voltage, high efficiency to 92.5 percent, low leakage current, low standby power, an internal EN 55032 Class B filter and an operating altitude of up to 5,000m. Built-in safeguards include overcurrent, over voltage and short circuit protection. MUI50 AC-DC power supplies are available in open-frame (3 x 1.5 x 1.08 inches), enclosed (3.65 x 1.91 x 1.56 inches) and pin-type (2.76 x 1.5 x 1.08 inches) formats, and they comply with RoHS and REACH regulations.
MUI50 Series Specification Sheet PDF
Daburn Part Number |
Input Voltage Range Vdc
|
Output Voltage Vdc |
Output Current Natural Convection A |
Maximum Output Power W |
Efficiency % |
Input Power at No Load mW |
Maximum Capacitor Load µF |
MUI50-S5B
|
85-264 |
5 |
8,000 |
40 |
90.5 |
50 |
16,000 |
MUI50-S7.5B
|
85-264 |
7.5 |
6,670 |
50 |
90.5 |
50 |
8,900 |
MUI50-S9B
|
85-264 |
9 |
5,560 |
50 |
90.5 |
50 |
6,200 |
MUI50-S12B
|
85-264 |
12 |
4,170 |
50 |
92.5 |
50 |
3,500 |
MUI50-S15B
|
85-264 |
15 |
3,340 |
50 |
92.5 |
50 |
2,300 |
MUI50-S18B
|
85-264 |
18 |
2,780 |
50 |
92.5 |
100 |
1,600 |
MUI50-S24B
|
85-264 |
24 |
2,085 |
50 |
92.5 |
100 |
870 |
MUI50-S36B
|
85-264 |
36 |
1,390 |
50 |
91.5 |
100 |
390 |
MUI50-S48B
|
85-264 |
48 |
1,045 |
50 |
91.5 |
100 |
220 |
MUI50-S53B
|
85-264 |
53 |
950 |
50 |
91.5 |
100 |
180 |
Input and Output Specifications:
INPUT
- Operating input voltage range:
- AC Input: 85 - 264Vdc
- DC Input: 120 - 370Vdc
- Input Frequency: AC Input: 47 - 63Hz, AC Input
- Input Current:
- 100 VAC and full load: 1.4A
- 240 VAC and full load: 0.8A
- Leakage Current: 264 Vac: 100µA
- Start-up time: 1000ms
- Rise time: 15ms
- Hold-up time: 115 Vac and full load: 12ms
- Input Inrush Current: 230 Vac: 60A
- Input Protection: Internal fuse: T3.5A/250Vac
OUTPUT
- Output Power
- 40 W, 5 Vout, Full Load
- 50 W, Others
- Output Peak Power
- 56 W, 5 Vout
- 65 W, 7.5 Vout
- 70 W, Others
- Output Peak Power Time: 5 s
- Output Peak Power Duty: 20%
- Average Operation Power: 70%, % of Full Load
- Initial Set Voltage Accuracy: 230Vac and Full Load: ±1.0%
- Line Regulation: Low Line to High Line at Full Load: ±0.2%
- Load Regulation:
- No Load to Full Load, 5 Vout: ±0.7%
- No Load to Full Load,Others: ±0.5%
- 10% Load to 90% Load, 5 Vout: ±0.6%
- 10% Load to 90% Load, Others: ±0.4%
- Voltage Adjustability
- Single output, 5 Vout, 7.5 Vout, 9 Vout: ±10%
- Others: -20-10%
- Minimum Load: 0%
- Ripple and Noise (Measured by 20MHz)
- With a 10µF/25V 1206 X7R MLCC, 5 Vout, 7.5 Vout, 9 Vout: 75mVp-p
- With a 10µF/25V 1206 X7R MLCC, 12 Vout 15 Vout, 18 Vout: 100mVp-p
- With a 1µF/50V 1206 X7R MLCC, 24 Vout, 36 Vout: 100mVp-p
- With a 0.1µF/100V 1206 X7R MLCC, 48 Vout 53 Vout: 100mVp-p
- Temperature Coefficient: ±0.02%/ºC
- Transient Response
- 3% Vout Max., Load step from 50 ~ 75% change at 2.5A/µs, peak deviation
- 300µs, Recovery time
- Over Voltage Protection: % of Vout(nom); Latch mode: 115 - 135
- Over Load Protection: % of Iout rated; Hiccup mode: 165%
- Short Circuit Protection: Continuous, automatic recovery
More Specifications:
GENERAL
- Isolation Voltage:
- Input to Output, 1 minute, 2MOPP insulation: 4,000Vac
- Input (Output) to F.G.: 2,500Vac
- Isolation Resistance: 500Vdc, 0.1GΩ
- Switching Frequency
- 230 Vac, 5 Vout: 70-95kHz
- 230 Vac, 7.5 Vout: 95-120kHz
- 230 Vac, Others: 110-135kHz
ENVIRONMENTAL
- Operating Ambient Temperature:
- With derating, Natural convection: -40-85ºC
- Storage Temperature Range: -40-85ºC
- Operating Altitude: 5000m
- Shock: IEC60068-2-27
- Vibration: IEC60068-2-6
- Relative Humidity: Non-condensing, 5% to 95% RH
PHYSICAL
- Design Meet Safety Standard: IEC/EN/ANSI/AAMI ES 60601-1, IEC/EN/UL 62368-1
- Dimensions:
- Open type: 3" × 1.5" × 1.08" (76.2 × 38.1 × 27.5 mm)
- Enclosed type: 3.65" × 1.91" × 1.56" (92.7 × 48.6 × 39.5 mm)
- Pin type: 2.76" × 1.5" × 1.08" (70 × 38.1 × 27.5 mm)
- Weight:
- Open type: 78g (2.75oz)
- Enclosed type: 175g (6.18oz)
- Pin type: 75g (2.65oz)
- MTBF: MIL-HDBK-217F, Full load, 1.487 x 106 hrs
EMC CHARACTERISTICS
- EMI:
- EN55011, EN55032, EN60601-1-2 and FCC Part 18/15
- Conducted: Class B
- Radiated: Class B
- Harmonic Currents: EN61000-3-2, Full load, Class A
- Voltage Flicker: EN61000-3-3
- EMS: EN55035 and EN60601-1-2
- ESD: EN61000-4-2, Perf. Criteria A
- Radiated Immunity: EN61000-4-3, 20V/m, Perf. Criteria A
- Fast Transient: EN61000-4-4, ±2kV, Perf. Criteria A
- Surge: EN61000-4-5, DM ±1kV and CM±2kV, Perf. Criteria A
- Conducted Immunity: EN61000-4-6, 20 Vr.m.s, Perf. Criteria A
- Power Frequency Magnetic Field: EN61000-4-8, 10 A/m, Perf. Criteria A
- Dip and Interruptions: EN61000-4-11
|
Product Reviews
|
|