The MUIP300 AC-DC power supply for medical applications delivers up to 300 watts and up to 360 watts peak power for 5 seconds (typ.) while accepting a universal input of 85 to 264V AC. It comes in four styles: open type (2.09 x 4.00 x 1.26 inches), base plate type (2.44 x 4.60 x 1.56 inches) and enclosed and DIN rail types, both measuring 2.44 x 4.6 x 2.32 inches. Features include adjustable output voltage, high efficiency up to 93 percent, 4,000V AC reinforced insulation, low standby power, low leakage current and an internal EN 55032 Class B filter. The Isolation Class I and Class II units also provide remote ON/OFF and fan speed control, a power good signal and an operating altitude of 5,000m. This medical AC-DC power supply offers 2xMOPP isolation for patient safety, meets IEC/EN/ANSI/AAMI ES 60601-1 and IEC/EN/UL 62368-1 safety standards and offers built-in overcurrent, over voltage, short circuit and over temperature protection. The MUIP300 complies with RoHS and REACH regulations.

MUIP300 Series Specification Sheet PDF

Daburn Part
Number

Input Voltage Range
Vdc

Output Voltage
Vdc
Output Current Natural Convection
A
Output Current Forced Air Cooling with 21 CFM
mA
Input Power at No Load
W
Efficiency
%
Maximum Capacitor Load
µF

MUIP300-12S

85-264 12 15 25 0.3 91 20,000

MUIP300-15S

85-264 15 12 20 0.3 92 12,000

MUIP300-18S

85-264 18 10 16.66 0.3 93 9,000

MUIP300-24S

85-264 24 7.5 12.5 0.3 93 2,400

MUIP300-28S

85-264 28 6.42 10.71 0.3 93 2,000

MUIP300-36S

85-264 36 5 8.33 0.3 93 1,000

MUIP300-48S

85-264 48 3.75 6.25 0.3 93 650

MUIP300-53S

85-264 53 3.4 5.67 0.3 93 470



Input and Output Specifications:


INPUT

  • Voltage Range:
    • AC Input: 85 - 264Vdc
    • DC Input: 120 - 370Vdc
  • Input Frequency: AC Input: 47 - 63Hz, AC Input
  • Input Current:
    • 100 VAC and full load: 3.9A
    • 240 VAC and full load: 1.6A
  • No Load Input Power
    • 230 Vac, Option F (with Fan): 3 W
    • Others: 0.3 W
  • Leakage Current: 264 Vac: 100µA
  • Power factor: 0.9
  • Start-up time: 2000ms
  • Rise time: 30ms
  • Hold-up time: 115 Vac and 225 W: 10ms
  • Input Inrush Current: 230 Vac: 70A
  • Input Protection: Internal fuse: T5.0A/250Vac

OUTPUT

  • Output Power
    • 300 W, Forced air cooling
    • 180 W, Natural convection
  • Output Peak Power: 360 W
  • Output Peak Power Time: 5 s
  • Output Peak Power Duty: 20%
  • Output Peak Power Average Operation: 50%, % of Full Load
  • Initial Set Voltage Accuracy: 230Vac and Full Load: ±1.0%
  • Line Regulation: Low Line to High Line at Full Load: ±0.2%
  • Load Regulation:
    • No Load to Full Load: ±0.5%
    • 10% to 90% Load: ±0.4%
  • Voltage Adjustability: ±10%
  • Minimum Load: 0%
  • Ripple and Noise (Measured by 20MHz bandwidth with a 1µF/100V 1206 X7R MLCC)
    • 12Vout: 120mVp-p
    • 15Vout: 150mVp-p
    • 18Vout: 180mVp-p
    • 24Vout: 240mVp-p
    • 28Vout: 280mVp-p
    • 36Vout: 360mVp-p
    • 48Vout: 480mVp-p
    • 53Vout: 530mVp-p
  • Temperature Coefficient: ±0.02%/ºC
  • Transient Response Peak Deviation: Load step from 50-75% change at 2.5A/μs: 3% Vout
  • Transient Response Recovery Time: Recovery within 1% Vout: 600μs
  • Over Voltage Protection: % of Vout(nom); Latch mode: 115 - 135
  • Over Load Protection: % of Iout rated; Hiccup mode: 150%
  • Short Circuit Protection: Continuous, automatic recovery
  • Main output mode remote control
    • Open or 3 ~ 12 VDC, Positive Logic, Main power ON
    • Short or 0 ~ 1.2VDC, Referenced to "-Control", Main power OFF
    • -0.5mA Min, 1mA Max, Standby power always present, Input current of control
  • Main output Power Good signal
    • Power good, Low (Referenced to "GND")
    • Power off, Open collector
  • Standby Power Supply: 1000mA, Standby and fan power supply total power 8W, 5Vout
  • Fan Power Supply: 500mA, 12Vout

More Specifications: 

 

GENERAL

  • Isolation Voltage:
    • Input to Output, 1 minute (Reinforced insulation): 4,000Vac
    • Input (Output) to F.G.: 2,500Vac
  • Isolation Resistance: 500Vdc, 0.1GΩ
  • Switching Frequency: 230Vac, Full load: 140kHz


ENVIRONMENTAL

  • Operating Ambient Temperature:
    • With derating, Option -F (with Fan): -40-80ºC
    • Others: -40-85ºC
  • Storage Temperature Range:
    • With derating, Option -F (with Fan): -40-80ºC
    • Others: -40-85ºC
  • Over temperature protection: 125°C (Internal thermistor; Hiccup mode)
  • Operating Altitude: 5000m
  • Shock: IEC60068-2-27
  • Vibration: IEC60068-2-6
  • Relative Humidity: Non-condensing, 5% to 95% RH

 

PHYSICAL

  • Design Meet Safety Standard: IEC/EN/ANSI/AAMI ES 60601-1, IEC/EN/UL 62368-1
  • Weight:
    • Open type: 210g (7.40oz)
    • Enclosed type: 318g (11.21oz)
    • Base plate type: 260g (9.17oz)
    • Din rail type: 340g (11.99oz)
  • MTBF: MIL-HDBK-217F Ta=25°C, Full load, 1.056 x 106 hrs
  • Dimensions:
    • Open type: 2.09" × 4" × 1.26"
    • Enclosed type: 2.44" × 4.6" × 2.32"
    • Base plate type: 2.44" × 4.6" × 1.56"
    • Din rail type: 2.44" × 4.6" × 2.32"

 

 EMC CHARACTERISTICS

  • EMI:
    • EN55011, EN55032, EN60601-1-2 and FCC Part 18/15
    • Conducted: Class B
    • Radiated: Class A
  • Harmonic Currents: EN61000-3-2, Full load, Class A
  • Voltage Flicker: EN61000-3-3
  • EMS: EN55035 and EN60601-1-2
  • ESD: EN61000-4-2, Perf. Criteria A
  • Radiated Immunity: EN61000-4-3, 20V/m, Perf. Criteria A
  • Fast Transient: EN61000-4-4, ±2kV, Perf. Criteria A
  • Surge: EN61000-4-5, DM ±1kV and CM±2kV, Perf. Criteria A
  • Conducted Immunity: EN61000-4-6, 20 Vr.m.s, Perf. Criteria A
  • Power Frequency Magnetic Field: EN61000-4-8, 30 A/m, Perf. Criteria A
  • Dip and Interruptions: EN61000-4-11

 