Polytron MUIPCP240 AC-DC Power Supply for Medical Applications
|
Price (Each): Call or e-mail for quotes or to order.
Available to Order
Item Number: MUIPCP240
|
MUIPCP240 Series Medical Power Supplies
MUIPCP240 Series medical power supplies come in a 5 x 3 x 1.32-inch open-frame package, accept a universal input of 85 to 264V AC, and they deliver 240 watts of output power at full load and 300 watts peak power. The Isolation Class I and Class II units provide 4,000V AC input to output 2xMOPP reinforced insulation for patient safety and satisfy IEC/EN/ANSI/AAMI ES 60601-1 and IEC/EN/UL 62368-1 safety standards. Additional features include an adjustable output voltage, high efficiency to 91 percent, low leakage current, remote ON/OFF, an internal EN 55032 Class B filter and an operating altitude of up to 5,000m. The units do not require a minimum load, and built-in safeguards include over temperature, overcurrent, overvoltage and short circuit protections. MUIPCP240 power supplies are compliant to RoHS and REACH regulations.
MUIPCP240 Series Specification Sheet PDF
Daburn Part Number |
Input Voltage Range Vdc
|
Output Voltage Vdc |
Output Current |
Output Power |
Efficiency % |
Maximum Capacitor Load µF |
Forced Air Cooling With 10 CFM @Full Load A |
Maximum Peak Power W |
@Full Load W |
Maximum Peak Power W |
MUIPCP240-S12
|
85-264 |
12 |
20 |
25 |
240 |
300 |
90 |
10,000 |
MUIPCP240-S15
|
85-264 |
15 |
16 |
20 |
240 |
300 |
90 |
6,800 |
MUIPCP240-S24
|
85-264 |
24 |
10 |
12.5 |
240 |
300 |
90 |
2,400 |
MUIPCP240-S28
|
85-264 |
28 |
8.58 |
10.72 |
240 |
300 |
90 |
2,000 |
MUIPCP240-S36
|
85-264 |
36 |
6.67 |
8.34 |
240 |
300 |
90 |
1,200 |
MUIPCP240-S48
|
85-264 |
48 |
5 |
6.25 |
240 |
300 |
91 |
680 |
MUIPCP240-S54
|
85-264 |
54 |
4.45 |
5.56 |
240 |
300 |
91 |
680 |
Input and Output Specifications:
INPUT
- Voltage Range:
- AC Input: 85 - 264Vdc
- DC Input: 88 - 370Vdc
- Input Frequency: AC Input: 47 - 63Hz, AC Input
- Input Current:
- 100 VAC and full load: 3.8A
- 240 VAC and full load: 1.5A
- No Load Input Power
- 230 Vac, Option -F3 (with Fan): 3 W
- Without fan: 0.3 W
- Leakage Current: 264 Vac: 100µA
- Power factor: 0.95
- Start-up time: 1,500ms
- Rise time: 20ms
- Hold-up time: 115 Vac and full load: 25ms
- Input Inrush Current: 230 Vac, cold start at 25°C: 80A
- Input Protection: Internal fuse: T5.0A/300Vac
OUTPUT
- Output Power
- 240 W, Forced air cooling
- 180 W, Natural convection
- Output Peak Power: 300 W
- Output Peak Power Time: 10 s
- Output Peak Power Duty: 20%
- Average Operation Power: 80%, % of Full Load
- Initial Set Voltage Accuracy: 230Vac and Full Load: ±1.0%
- Line Regulation: Low Line to High Line at Full Load: ±0.2%
- Load Regulation:
- No Load to Full Load: ±0.5%
- 10% to 90% Load: ±0.4%
- Voltage Adjustability
- 12~48 Vout: ±10%
- 54 Vout: -20% Min, 0% Max
- Minimum Load: 0%
- Ripple and Noise (Measured by 20MHz bandwidth)
- 12Vout: 120mVp-p, with a 10µF/25V 1206 X7R MLCC
- 15Vout: 150mVp-p, with a 10µF/25V 1206 X7R MLCC
- 24Vout: 240mVp-p, with a 1µF/50V 1206 X7R MLCC
- 28Vout: 280mVp-p, with a 1µF/50V 1206 X7R MLCC
- 36Vout: 360mVp-p, with a 1µF/50V 1206 X7R MLCC
- 48Vout: 480mVp-p, with a 1µF/100V 1206 X7R MLCC
- 54Vout: 540mVp-p, with a 1µF/100V 1206 X7R MLCC
- Temperature Coefficient: ±0.02%/ºC
- Transient Response
- 3% Vout, Load step from 100~75% change at 2.5 A/µs, peak deviation
- 1,000 µs, recovery time
- Over Voltage Protection
- 130% of Vout(nom); 12~36 Vout; LAtch mode
- 59.5% of Vout(nom); 48~54 Vout; Latch mode
- Over Load Protection: % of Iout rated; Hiccup mode: 180%
- Remote ON/OFF (Option -L suffix)
- 0~0.8 VDC or Open, External power supply is required between +Ctrl and -Ctrl, Output OFF
- 4.5~12.5 VDC, Output ON
- 20 Max, mA, Input current
- Fan power supply: 12V / 300mA, Fixed fan speed function
- Short circuit protection: Continuous, automatic recovery
More Specifications:
GENERAL
- Isolation Voltage:
- Input to Output, 1 minute, 2MOPP insulation: 4,000Vac
- Input (Output) to F.G.: 2,500Vac
- Isolation Resistance
- 500Vdc: 0.1GΩ
- 230 Vac, Full load: 60GΩ
ENVIRONMENTAL
- Operating Ambient Temperature, With derating: -40-85ºC
- Storage Temperature Range: -40-85ºC
- Operating Altitude: 5,000m
- Thermal Shock: MIL-STD-810F
- Shock: IEC60068-2-27
- Vibration: IEC60068-2-6
- Relative Humidity: Non-condensing, 5% to 95% RH
PHYSICAL
- Design Meet Safety Standard: IEC/EN/ANSI/AAMI ES 60601-1, IEC/EN/UL 62368-1
- Dimensions: 5" × 3" × 1.32" (127 × 76.2 × 33.6mm)
- Weight: 340g (11.99oz)
- MTBF: 7.252 × 105hrs, MIL-HDBK-217F Ta=25°C, Full load
EMC CHARACTERISTICS
- EMI:
- EN55011, EN55032, EN60601-1-2 and FCC Part 18/15
- Conducted: Class B
- Radiated: Class B
- Harmonic Currents: EN61000-3-2, Full load, Class D
- Voltage Flicker: EN61000-3-3
- EMS: EN55035 and EN60601-1-2
- ESD: EN61000-4-2, Perf. Criteria A
- Radiated Immunity: EN61000-4-3, 20V/m, Perf. Criteria A
- Fast Transient: EN61000-4-4, ±2kV, Perf. Criteria A
- Surge: EN61000-4-5, DM ±1kV and CM±2kV, Perf. Criteria A
- Conducted Immunity: EN61000-4-6, 20 Vr.m.s, Perf. Criteria A
- Power Frequency Magnetic Field: EN61000-4-8, 30 A/m, Perf. Criteria A
- Dip and Interruptions: EN61000-4-11
|
Product Reviews
|
|