Polytron MUIPP150 AC-DC Power Supply for Medical Applications
|
Price (Each): Call or e-mail for quotes or to order.
Available to Order
Item Number: MUIPP150
|
MUIPP150 AC-DC Power Supply for Medical Applications
The MUIPP150 AC-DC power supply for medical applications delivers up to 150 watts and up to 200 watts peak power for 5 seconds (typ.) while accepting a universal input of 85 to 264V AC. The unit is available in open frame enclosed and DIN rail configurations with a compact 3” x 2” package size. Features include adjustable output voltage, high efficiency up to 94 percent, 4,000V AC reinforced insulation, low standby power, low leakage current and an internal EN 55032 Class B filter. The Isolation Class I and Class II MUIPP150 also provides overcurrent, overvoltage and over temperature protections. The operating altitude is 5,000m. This medical AC-DC power supply offers 2xMOPP isolation for patient safety, meets IEC/EN/ANSI/AAMI ES 60601-1 and IEC/EN/UL 62368-1 safety standards, and complies with RoHS and REACH regulations.
MUIPP150 Series Specification Sheet PDF
Daburn Part Number |
Input Voltage Range Vac
|
Output Voltage Vdc |
Output Current Natural Convection A |
Output Current Forced Air Cooling with 10 CFM A |
Efficiency % |
Input Power @ No Load mW |
Maximum Capacitor Load µF |
MUIPP150-12S
|
85-264 |
12 |
10.84 |
12.5 |
92 |
0.2 |
8600 |
MUIPP150-15S
|
85-264 |
15 |
8.67 |
10 |
92 |
0.2 |
5400 |
MUIPP150-18S
|
85-264 |
18 |
7.23 |
8.34 |
92 |
0.2 |
4000 |
MUIPP150-24S
|
85-264 |
24 |
5.42 |
6.25 |
93 |
0.2 |
2200 |
MUIPP150-28S
|
85-264 |
28 |
4.65 |
5.36 |
93 |
0.2 |
1500 |
MUIPP150-36S
|
85-264 |
36 |
3.62 |
4.17 |
93 |
0.2 |
1000 |
MUIPP150-48S
|
85-264 |
48 |
2.71 |
3.13 |
94 |
0.2 |
560 |
MUIP300-54S
|
85-264 |
54 |
2.42 |
2.79 |
94 |
0.2 |
470 |
Input and Output Specifications:
INPUT
- Voltage Range:
- AC Input: 85 - 264Vdc
- DC Input: 88 - 370Vdc
- Input Frequency: AC Input: 47 - 63Hz, AC Input
- Input Current:
- 100 VAC and at 150W full load: 3A
- 240 VAC and at 150W full load: 1.5A
- No Load Input Power
- Leakage Current: 264 Vac: 100µA
- Power factor: 0.95
- Start-up time: 1500ms
- Rise time: 15ms
- Hold-up time: 115 Vac and 130 W: 10ms
- Input Inrush Current: 230 Vac: 80A
- Input Protection: Internal fuse: T4.0A/250Vac
OUTPUT
- Output Power
- 150 W, Forced air cooling with 10CFM
- 130 W, Natural convection
- Output Peak Power: 200 W
- Output Peak Power Time: 10 s
- Output Peak Power Duty: 20%
- Average Operation Power: 50%, % of Full Load
- Initial Set Voltage Accuracy: 230Vac and Full Load: ±1.0%
- Line Regulation: Low Line to High Line at Full Load: ±0.2%
- Load Regulation:
- No Load to Full Load: ±0.5%
- 10% to 90% Load: ±0.4%
- Voltage Adjustability: ±7%
- Minimum Load: 0%
- Ripple and Noise (Measured by 20MHz bandwidth)
- 12Vout: 120mVp-p with a 1µF/25V 1206 X7R MLCC
- 15Vout: 150mVp-p with a 1µF/25V 1206 X7R MLCC
- 18Vout: 180mVp-p with a 1µF/25V 1206 X7R MLCC
- 24Vout: 240mVp-p with a 1µF/50V 1206 X7R MLCC
- 28Vout: 280mVp-p with a 1µF/50V 1206 X7R MLCC
- 36Vout, 48Vout, 54Vout: 360mVp-p with a 1µF/100V 1206 X7R MLCC
- Temperature Coefficient: ±0.02%/ºC
- Transient Response Peak Deviation: Load step from 75-50% change at 2.5A/μs: 3% Vout
- Transient Response Recovery Time: 600μs
- Over Voltage Protection: % of Vout(nom); Latch mode: 110 - 135%
- Over Load Protection: % of Iout rated; Hiccup mode: 160%
- Short Circuit Protection: Continuous, automatic recovery
More Specifications:
GENERAL
- Isolation Voltage:
- Input to Output, 1 minute, 2MOPP insulation: 4,000Vac
- Input (Output) to F.G.: 2,500Vac
- Isolation Resistance: 500Vdc, 0.1GΩ
- Switching Frequency: 230Vac, Full load: 100kHz
ENVIRONMENTAL
- Operating Ambient Temperature:
- Storage Temperature Range:
- Over temperature protection: 125°C (Internal thermistor; Hiccup mode)
- Operating Altitude: 5000m
- Shock: IEC60068-2-27
- Vibration: IEC60068-2-6
- Relative Humidity: Non-condensing, 5% to 95% RH
PHYSICAL
- Design Meet Safety Standard: IEC/EN/ANSI/AAMI ES 60601-1, IEC/EN/UL 62368-1
- Weight:
- Open type: 134g (4.70oz)
- Enclosed type: 190g (6.70oz)
- Din rail type: 212g (7.48oz)
- MTBF: MIL-HDBK-217F Ta=25°C, Full load, 7.245 x 105 hrs
EMC CHARACTERISTICS
- EMI:
- EN55011, EN55032, EN60601-1-2 and FCC Part 18/15
- Conducted: Class B
- Harmonic Currents: EN61000-3-2, Full load, Class D
- Voltage Flicker: EN61000-3-3
- EMS: EN55035 and EN60601-1-2
- ESD: EN61000-4-2, Perf. Criteria A
- Radiated Immunity: EN61000-4-3, 20V/m, Perf. Criteria A
- Fast Transient: EN61000-4-4, ±2kV, Perf. Criteria A
- Surge: EN61000-4-5, DM ±1kV and CM±2kV, Perf. Criteria A
- Conducted Immunity: EN61000-4-6, 20 Vr.m.s, Perf. Criteria A
- Power Frequency Magnetic Field: EN61000-4-8, 30 A/m, Perf. Criteria A
- Dip and Interruptions: EN61000-4-11
|